160 x 96 (or 192 x 64 depending on the specific sub-model) graphic LCD. Printing System: High-speed thermal array (8 dots/mm).
Feed a 1mV signal from the simulator. The device should produce a 10mm deflection. If it does not, adjustment via the internal potentiometer or service menu software is required. carewell ecg 1101 service manual verified
Ensure the internal ribbon cable connecting the ECG input board to the mainboard is seated correctly. Printing Issues (Faint or Blank) 160 x 96 (or 192 x 64 depending
Complete electrical safety testing (Leakage current should be <10µA). Conclusion carewell ecg 1101 service manual verified
AC 100V–240V (50/60Hz) or DC 12V rechargeable battery. 2. Common Troubleshooting Codes and Errors