Never plunge the device into water or use soaking wet cloths. Troubleshooting & Technical Servicing
Regular servicing prevents critical failure during emergencies. Users are advised to perform frequent visual and functional checks. To avoid the "memory effect" and prolong battery life, the
If the unit cannot deliver 15 shocks after a full charge, the accumulator should be replaced by an authorized service department. 2. Visual Inspection Checklist primedic defi b service manual work
Verifying output at 10J, 50J, 100J, 200J, 300J, and 360J (± 15%).
Confirming the "R" wave is detected correctly for cardioversion. Never plunge the device into water or use soaking wet cloths
Annual or biennial technical safety controls are recommended. These often include:
A fully charged unit must be capable of delivering at least 15 shocks at maximum energy (360 Joules). Charge Time: Approximately 3.5 hours for a full charge. Essential Maintenance & Care To avoid the "memory effect" and prolong battery
14.4V, 1400mA battery pack consisting of 12 parallel cells.
Use standard domestic cleaning agents and a clean, dry cloth.
Inspect the paddle cable insulation for fraying or damage.